First Use of Mini Gamma Cameras for Intra-operative Robotic SPECT ReconstructionPhilipp Matthies1, Kanishka Sharma1, Asl 1Computer Aided Medical Procedures (CAMP), Technische Universität München, Germany 2Department of Nuclear Medicine, Klinikum Rechts der Isar, Technische Universität München, Germany 3Institute of Biomathematics and Biometry Helmholtz Zentrum München, Germany Abstract. Different types of nuclear imaging systems have been used in the past, starting with pre-operative gantry-based SPECT systems and gamma cameras for 2D imaging of radioactive distributions. The main applications are concentrated on diagnostic imaging, since traditional SPECT systems and gamma cameras are bulky and heavy. With the development of compact gamma cameras with good resolution and high sensitivity, it is now possible to use them without a fixed imaging gantry. Mounting the camera onto a robot arm solves the weight issue, while also providing a highly repeatable and reliable acquisition platform. In this work we introduce a novel robotic setup performing scans with a mini gamma camera, along with the required calibration steps, and show the first SPECT reconstructions. The results are extremely promising, both in terms of image quality as well as reproducibility. In our experiments, the novel setup outperformed a commercial fhSPECT system, reaching accuracies comparable to state-of-the-art SPECT systems. LNCS 8149, p. 163 ff. lncs@springer.com
|